The effect of wear on ultrasonic scaler tip displacement amplitude
Resumen
AIM: During clinical usage, scaler tips may become worn and reduced in length. It is unknown what effect wear has on the magnitude of scaler tip vibrations when they are utilized under typical clinical loads. The aim of this investigation was to assess the effect of simulated wear on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer. MATERIALS AND METHODS: A Cavitron SPS (Dentsply) ultrasonic generator and three scaler insert designs (FSI-100, FSI-1000 and FSI-SLI-10S) were selected for the investigation. Tip vibration displacement amplitude was assessed unloaded and then contacting against tooth surfaces with loads of 0.5 and 1.0 N. Tips were then ground down by 1 mm and then 2 mm and scans were repeated.RESULTS: For all tips, load and length were found to be significant variables (p
Descriptores DECS
TECNICA ODONTOLOGICA DE ALTA VELOCIDAD; INSTRUMENTOS DENTALES; RASPADO DENTAL^sInstrumentación; ANALISIS DEL ESTRES DENTAL; DISEÑO DE EQUIPO; EQUIPO REUTILIZADO; RAYOS LASER; OSCILOMETRIA; TERAPIA POR ULTRASONIDO^sInstrumentación; VIBRACION